REPAIRING AND NOT THROWING AWAY

Richtige Fernseher haben Röhren!

Richtige Fernseher haben Röhren!

In Brief: On this site you will find pictures and technical information about Service Modes, Circuit Diagrams, Firmware Update procedure, Disassemble procedure, Universal remote control set-up codes, Troubleshooting and more....

If you go into the profession, you will obtain or have access to a variety of tech tips databases HERE IT IS Master Electronics Repair !.

These are an excellent investment where the saying: 'time-is-money' rules. However, to learn, you need to develop a general troubleshooting approach - a logical, methodical, method of narrowing down the problem. A tech tip database might suggest: 'Replace C536' for a particular symptom. This is good advice for a specific problem on one model. However, what you really want to understand is why C536 was the cause and how to pinpoint the culprit in general even if you don't have a service manual or schematic and your tech tip database doesn't have an entry for your sick TV or VCR.

While schematics are nice, you won't always have them or be able to justify the purchase for a one-of repair. Therefore, in many cases, some reverse engineering will be necessary. The time will be well spent since even if you don't see another instance of the same model in your entire lifetime, you will have learned something in the process that can be applied to other equipment problems.
As always, when you get stuck, checking out a tech-tips database may quickly identify your problem and solution.In that case, you can greatly simplify your troubleshooting or at least confirm a diagnosis before ordering parts.

Happy repairing!
Today, the West is headed for the abyss. For the ultimate fate of our disposable society is for that society itself to be disposed of. And this will happen sooner, rather than later.

How to use the site:

- If you landed here via any Search Engine, you will get what you searched for and you can search more using the search this blog feature provided by Google. You can visit more posts scrolling the left blog archive of all posts of the month/year,
or you can click on the main photo-page to start from the main page. Doing so it starts from the most recent post to the older post simple clicking on the Older Post button on the bottom of each page after reading , post after post.

You can even visit all posts, time to time, when reaching the bottom end of each page and click on the Older Post button.

- If you arrived here at the main page via bookmark you can visit all the site scrolling the left blog archive of all posts of the month/year pointing were you want , or more simple You can even visit all blog posts, from newer to older, clicking at the end of each bottom page on the Older Post button.
So you can see all the blog/site content surfing all pages in it.

- The search this blog feature provided by Google is a real search engine. If you're pointing particular things it will search IT for you; or you can place a brand name in the search query at your choice and visit all results page by page. It's useful since the content of the site is very large.

Note that if you don't find what you searched for, try it after a period of time; the site is a never ending job !

..............The bitterness of poor quality is remembered long after the sweetness of todays funny gadgets low price has faded from memory........ . . . . . .....
Don't forget the past, the end of the world is upon us! Pretty soon it will all turn to dust!

©2010, 2011, 2012, 2013, 2014 Frank Sharp - You do not have permission to copy photos and words from this blog, and any content may be never used it for auctions or commercial purposes, however feel free to post anything you see here with a courtesy link back, btw a link to the original post here , is mandatory.
All posts are presented here for informative, historical and educative purposes as applicable within fair use. NOTHING HERE IS FOR SALE !

Monday, 1 April 2019

TERUMO TE331 - TE332 -TERUFUSION SYRINGE PUMP - E2PROM INITIALIZE AND AGING

TERUMO TE331 - TE332 -Terufusion Syringe Pump - E2PROM INITIALIZE and AGING

TERUMO TE-331 – TE332 Terufusion Syringe Pump. How to initialize E2PROM.
Medical (Electronic) Equipment.
E2PROM INITIALIZING
RAM data is erased and E2PROM is refreshed by ROM data [default data at factory].  This operation resets all the settings to the default.  Readjust required.
  1. Set the internal DIP switches 1 and 2 to ON and OFF respectively.
  2. While holding down the [STOP] switch, press the On/Off switch.  The display first shows “888.8”, then “FACt”, and a short beep follows.  Release the switches and check the display has changed to “”dIAg”; which stands for Diagnosis.  Note:  If you press the On/Off switch without pressing the Stop switch “Er91” appears on the 7 segment display.  Turn off the pump, and repeat the procedure above.  If you do not release the switches at the beep, the display changes to ‘Er 8”.  Once ‘Er 8’ appears, you have to turn off the pump, and repeat the procedure above.
  3. When “dIAg” is displayed, press the Display Select switch  3 times.  Check the display shows “CLEA” which stands for clear.
  4. Press ‘Stop ‘switch, and the display shows “E or J”.  E is for Europe, and J is for Japan.
  5. Press and hold 'PURGE’ switch about 1.5 seconds to choose ‘E for Europe’.
  6. Initialization of E2PROM starts.  The display changes to ‘Eur’.  As the initialization proceeds, first “      “, then “00     “,and “000 “, lastly “0000” appears on the display.the switch for Japan is the Start.  If the Start switch is pressed, “JP” appears on the display.
  7. When successfully completed, can hear a short beep and see the display of “good”.
  8. In case of failure, “bAd” is displayed.  Then the display returns to “CLEA”.
  9. Press On/Off switch to Exit the initializing E2PROM mode.
Aging.
  1. Set the DIP switches 1 and 2 to On and Off, respectively.
  2. While holding the Stop switch, press On/Off switch.  The display first shows “888.8”, then “FACt” and a short beep follows.  Release the switches and check the display has changed to “dLAg” which stands for Diagnosis.  Note: If you press the On/Off switch without pressing the ‘Stop’ switch “Er91” appears on the 7 segment display.  Turn off the pump, and repeat the procedure above.  If you do not release the switches at the beep, the display changes to “Er 8”.  Once “Er 8” appears, you have to turn Off the pump and repeat the procedure above.
  3. When “dIAG” is displayed, press the Display Select switch 4 times.  Check the display shows “Aging”, which stands for Aging.
  4. Press Start switch, and Aging starts.
  5. When completed, press the On/Off switch to Exit the Aging Mode.
The following actions should be recognized while aging is performed.
* Motor rotates [slider moves approximately 9mm/h].
* All LED elements connected flash at 1 Hz / 50% duty.
* While the motor is rotating the minimum purge sound comes out.
Attention You can stop aging at any time by pressing the ‘Stop’ switch, and start again by pressing the Start switch.
* The motor stops if occlusion is detected with the load on the slider at approx. 4606N=4.7kgf.