REPAIRING AND NOT THROWING AWAY

Richtige Fernseher haben Röhren!

Richtige Fernseher haben Röhren!

In Brief: On this site you will find pictures and technical information about Service Modes, Circuit Diagrams, Firmware Update procedure, Disassemble procedure, Universal remote control set-up codes, Troubleshooting and more....

If you go into the profession, you will obtain or have access to a variety of tech tips databases HERE IT IS Master Electronics Repair !.

These are an excellent investment where the saying: 'time-is-money' rules. However, to learn, you need to develop a general troubleshooting approach - a logical, methodical, method of narrowing down the problem. A tech tip database might suggest: 'Replace C536' for a particular symptom. This is good advice for a specific problem on one model. However, what you really want to understand is why C536 was the cause and how to pinpoint the culprit in general even if you don't have a service manual or schematic and your tech tip database doesn't have an entry for your sick TV or VCR.

While schematics are nice, you won't always have them or be able to justify the purchase for a one-of repair. Therefore, in many cases, some reverse engineering will be necessary. The time will be well spent since even if you don't see another instance of the same model in your entire lifetime, you will have learned something in the process that can be applied to other equipment problems.
As always, when you get stuck, checking out a tech-tips database may quickly identify your problem and solution.In that case, you can greatly simplify your troubleshooting or at least confirm a diagnosis before ordering parts.

Happy repairing!
Today, the West is headed for the abyss. For the ultimate fate of our disposable society is for that society itself to be disposed of. And this will happen sooner, rather than later.

How to use the site:

- If you landed here via any Search Engine, you will get what you searched for and you can search more using the search this blog feature provided by Google. You can visit more posts scrolling the left blog archive of all posts of the month/year,
or you can click on the main photo-page to start from the main page. Doing so it starts from the most recent post to the older post simple clicking on the Older Post button on the bottom of each page after reading , post after post.

You can even visit all posts, time to time, when reaching the bottom end of each page and click on the Older Post button.

- If you arrived here at the main page via bookmark you can visit all the site scrolling the left blog archive of all posts of the month/year pointing were you want , or more simple You can even visit all blog posts, from newer to older, clicking at the end of each bottom page on the Older Post button.
So you can see all the blog/site content surfing all pages in it.

- The search this blog feature provided by Google is a real search engine. If you're pointing particular things it will search IT for you; or you can place a brand name in the search query at your choice and visit all results page by page. It's useful since the content of the site is very large.

Note that if you don't find what you searched for, try it after a period of time; the site is a never ending job !

..............The bitterness of poor quality is remembered long after the sweetness of todays funny gadgets low price has faded from memory........ . . . . . .....
Don't forget the past, the end of the world is upon us! Pretty soon it will all turn to dust!

©2010, 2011, 2012, 2013, 2014 Frank Sharp - You do not have permission to copy photos and words from this blog, and any content may be never used it for auctions or commercial purposes, however feel free to post anything you see here with a courtesy link back, btw a link to the original post here , is mandatory.
All posts are presented here for informative, historical and educative purposes as applicable within fair use. NOTHING HERE IS FOR SALE !

Friday, 8 February 2019

HOW TO CALIBRATE VELA VENTILATOR

How to calibrate Vela ventilator


Operational Verification and Calibration
Prior to using the Vela ventilator on a new patient the following checks should be carried out to ensure optimum performance. Verification testing should always be performed “off patient”.
Note: If any portion of the following performance check fails, and you are unable to correct the problem, contact your VIASYS Healthcare Certified Service Technician.
First perform the User Verification Tests (UVTs). To access the UVTs, do the following’
1. After disconnecting the patient, turn the ventilator OFF (i.e., STANDBY).
2. Press and hold the Accept button.
3. While holding the Accept button, turn the ventilator ON. Continue to hold the button until the ventilator completes the Power On Self Tests (POST).
4. Release the Accept button when the UVT menu appears in the screen. The Audible Alarm sounds. Press Alarm Reset button to clear.
5. Press the Patient Removed touch screen icon, the UVT test selection screen displays.
6. Press the appropriate touch screen icon to begin a test.
Lamp Test
Run this test to check the front lamps to make sure they are functioning properly.
1. Press the LampTest touch screen icon to start the test. The ventilator illuminates all front panel LEDs.
2. Press the LampTest touch screen icon again to turn the LEDs off and exit the test.
Switch Test
Run this test to check the front panel membrane switches to make sure they are working properly.
1. Press the SWITCH TEST icon.
2. Press each membrane switch control in turn. Watch for the name of the control to appear in the message bar at the bottom left of the touch screen as follows:
3. Press the SWITCH TEST icon again to exit the test.
Alarm Test
Run this test to check the audible alarm.
1. Press the Alarm Test touch screen icon to start the test. The audible alarm sounds.
2. Press the Alarm Test touch screen icon again to silence the audible alarm and exit the test.
Filter Test
Run this test to check the differential pressure across the turbine inlet filter. A high differential pressure may indicate a dirty or occluded filter.
1. Press the Filter Test touch screen icon. The turbine accelerates flow to 140 lpm and checks the differential pressure to make sure it is within range. At the end of the test, the ventilator displays a pass or fail message. If the test passes, the following message appears:
x.x Passed
where x.x is the average turbine differential pressure and P indicates the test passed. If the test fails, the following message appears:
x.x Failed
where x.x is the average turbine differential pressure and F indicates the test failed:
2. If the test fails, check the rotation of the turbine inlet filter to make sure that the seam is horizontally oriented and run the test again. If the test fails again, contact a Bird Certified Service Technician.
Leak Test
Note:This test should be performed with all circuit accessories installed (e.g., humidifier, water traps, and so on.) Make sure al connections are secure and all openings occluded before beginning the test.
Run this test to make sure the patient breathing circuit is not leaking.
1. Attach a one-liter test lung (P/N 33754) at the patient breathing circuit wye.
2. Press the Leak Test touch screen icon to run the test. The test begins by increasing the pressure in the patient breathing circuit to 60 cmH2O. The ventilator then displays the following messages in sequence:
Leak test requested
Leak test in progress
3. The ventilator then waits 10 seconds and measures the circuit pressure again. If the difference between the starting and ending measurements is less than or equal to 6 cmH2O, the test passes and the ventilator displays the following message:
xx.x Passed
where xx.x is the ending measurement.
4. Otherwise, if the difference is more than 6 cmH2O, the test fails and the ventilator displays the following message:
xx.x Failed
5. If the test fails, check all connections to make sure there are no leaks and repeat the test.
Exit
To exit the UVTs press the EXIT touch screen icon.